inforMED
MalfunctionCAD

TEC

Received Jul 20, 2026 · Event occurred Jun 22, 2026

Report 2112667-2026-04575 · MDR key 25882776

Device

Generic name

Vaporizer

Manufacturer

Datex-ohmeda, Inc.

Model number

820 ISO

Catalog number

NA

Lot number

NA

Product problems

  • Output below Specifications

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT AGENT DELIVERY LESS THAT -20% OF SETTING. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

A GE HEALTHCARE¿S INVESTIGATION INTO THE REPORTED OCCURRENCE IS ONGOING. A FOLLOW-UP REPORT WILL BE ISSUED WHEN THE INVESTIGATION HAS BEEN COMPLETED. NO REPAIR INFORMATION AVAILABLE. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS MADISON - 3030 OHMEDA DR, USA MADISON, WI 53718.