inforMED
InjuryJCW

AMS 700 LGX? TENACIO? PUMP

Received Jul 20, 2026 · Event occurred Jun 22, 2026

Report 2124215-2026-38778 · MDR key 25876879

Device

Generic name

Prosthesis, Penis, Inflatable

Model number

72404456

Catalog number

72404456

Lot number

1100701255

Product problems

  • Deflation Problem
  • Inflation Problem
  • Material Deformation

Patient

75 YR · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE PATIENT WITH THIS INFLATABLE PENILE PROSTHESIS (IPP) WAS UNABLE TO PROPERLY INFLATE AND DEFLATE THE DEVICE PUMP MECHANISM. UPON INSPECTION, A KINK IN THE TUBING BETWEEN THE PUMP AND THE RESERVOIR WAS IDENTIFIED. AS A RESULT, THE CYLINDERS WERE EXPLANTED, AND THE EXISTING RESERVOIR WAS DRAINED AND RETAINED. A COMPLETE NEW DEVICE SYSTEM WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED.

Additional Manufacturer Narrative

MODEL NUMBER/CATALOG NUMBER: 720185-01. SERIAL NUMBER: N/A. BATCH/LOT NUMBER: 1100685087. MODEL/CATALOG DESCRIPTION: RESERVOIR FLAT IZ 100 ML. UNIQUE IDENTIFIER (UDI) #: (B)(4).