OR1 NEO IP, ENH2 2.5
Received Jul 17, 2026
Report 9610617-2026-01656 · MDR key 25860691
Device
Generic name
Or1 Neo Ip, Enh2 2.5
Manufacturer
Karl Storz Se & Co. KgModel number
WUIS2548-01Catalog number
WUIS2548-01
Product problems
- Fire
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
ACCORDING TO THE INFORMATION RECEIVED, A SMALL FIRE THAT OCCURRED AT PIH HEALTH WHITTIER, OR 11 LAST NIGHT THAT WAS TRACED BACK TO THE CABINET HOUSING ALL THE KARL STORZ INTEGRATION EQUIPMENT. AT THIS TIME, THE FACILITY HAS SUSPENDED ALL CASES IN EVERY OR SUITE UNTIL KARL STORZ CAN ASSESS EACH ROOM AND PROVIDE CLEARANCE.
Additional Manufacturer Narrative
THE DEVICE WILL NOT BE FORWARDED TO THE MANUFACTURING SITE FOR INVESTIGATION. RATHER, A SERVICE TECH WILL BE SENT TO EVALUATE THE DEVICE IN THE FIELD. SHOULD RELEVANT ADDITIONAL INFORMATION / INVESTIGATION RESULTS BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED UNSOLICITED. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID: (B)(4). THIS PRODUCT IS MANUFACTURED AT KARL STORZ ENDOSCOPY AMERICA, 13803 N PROMENADE BLVD, STAFFORD, TX 77477; HOWEVER, IT IS DESIGNED IN GERMANY.