MalfunctionJJG

PICCOLO XPRESS CHEMISTRY ANALYZER

Received Jul 17, 2026 · Event occurred Jun 24, 2026

Report 2939693-2026-00017 · MDR key 25857143

Device

Generic name

Piccolo Xpress Chemistry Analyzer

Manufacturer

Abaxis, Inc.

Catalog number

1100-0000

Product problems

  • High Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE PATIENT WAS NOT TREATED BASED ON THE PICCOLO ANALYZER RESULTS. A SAMPLE WAS SENT TO A THIRD-PARTY LABORATORY FOR CONFIRMATION. THE END USER PREFERRED NOT TO DISCLOSE PATIENT HISTORY. ABAXIS TECHNICAL SUPPORT PERFORMED TROUBLESHOOTING WITH THE CUSTOMER. THE CUSTOMER WAS ADVISED TO REPEAT QC; VERIFY REAGENT/STORAGE CONDITIONS AND EXPIRATION DATES; CONFIRM ENVIRONMENTAL CONDITIONS WERE WITHIN SPECIFICATIONS; AND ENSURE PROPER SPECIMEN HANDLING. INFORMATION REGARDING THE PATIENT¿S PRE-EXISTING CONDITIONS AND/OR CURRENT MEDICATIONS IS UNKNOWN. THE CLINIC WAS PROVIDED WITH MATERIALS TO PERFORM A LINEARITY CHECK AND WAS OFFERED THE OPTION TO SEND THE ANALYZER IN FOR SERVICE. IT IS UNKNOWN WHETHER THE END USER HAS ELECTED TO SEND THE INSTRUMENT IN FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE ISSUED UPON COMPLETION OF THE INVESTIGATION. H3 OTHER TEXT : AWAITING CUSTOMERS DECISION TO RETURN

Description of Event or Problem

A HEALTH CARE PROFESSIONAL REPORTED AN UNEXPECTEDLY HIGH POTASSIUM (K+) RESULT USING THE PICCOLO XPRESS ANALYZER AND THE PICCOLO COMPREHENSIVE METABOLIC PANEL (CMP), LOT#6042FC1 ERROR CODES: CHEM K+ / K+ PROBLEM OR QUESTION.