inforMED
MalfunctionMCW

VIPERWIRE ADVANCED WITH FLEX TIP

Received Jul 16, 2026 · Event occurred Jul 8, 2026

Report MW5190805 · MDR key 25842665

Device

Generic name

Catheter, Peripheral, Atherectomy

Lot number

592016-1

Patient

86 YR · Male

  • Foreign Body In Patient

Narrative

Description of Event or Problem

PATIENT IN CATH LAB FOR LOWER EXTREMITY ANGIOGRAM AND INTERVENTION. LEFT POSTERIOR TIBIAL ARTERY WIRED WITH .014 X 300 COMMAND WIRE AND .018 CXI MICROCATHETER. COMMAND WIRE EXCHANGED WITH .014 X 335 CM VIPERWIRE FOR ATHERECTOMY. ATHERECTOMY PERFORMED BY 1.25 X 145 CM CSI MICRO DIAMONDBACK CATHETER. CSI MICROCATHETER REMOVED SUCCESSFULLY HOWEVER, VIPERWIRE BROKE OFF IN PATIENT'S BODY (LEFT FOOT). ATTEMPTS TO SNARE WIRE FRAGMENT MADE WITH 4 MM MICROSNARE (UNSUCCESSFUL). SUCCESSFULLY SNARED WITH 4-8 MM ENSNARE.