EVIS LUCERA COLONOVIDEOSCOPE

Received Jul 16, 2026 · Event occurred May 21, 2026

Report 9610595-2026-57220 · MDR key 25842125

Device

Generic name

Colonovideoscope

Model number

PCF-Q260AZI

Product problems

  • Overheating of Device

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: THE MOST PROBABLE CAUSE OF THE MALFUNCTION IS TRACED TO COMPONENT FAILURE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem

THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT BURN C-COVER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.