MalfunctionPZP

HYDROS ROBOTIC SYSTEM

Received Jul 16, 2026 · Event occurred Jun 19, 2026

Report 3012977056-2026-00148 · MDR key 25841930

Device

Generic name

Fluid Jet Removal System

Catalog number

HY1000T

Product problems

  • Visual Prompts will not Clear

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

A MALE PATIENT UNDERWENT AQUABLATION THERAPY TO TREAT SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT DURING AQUABLATION THERAPY, THE WATER JET POWER LEVEL CAME OUT AT A SIGNIFICANT LEVEL, SOUNDING MORE LIKE 50-100%. THERE WAS AN ERROR CODE (E005 - HIGH PRESSURE PUMP ERROR) AND AN ERROR (E433 COMMUNICATION ERROR) POPULATED ON SCREEN, REQUESTING A REBOOT. WHEN THE TREATING SURGEON WENT IN WITH THE CYSTOSCOPE TO VISUALIZE THE BLADDER FLOOR, FLUFFY WHITE TISSUE WAS NOTED ON THE TRIGONE BETWEEN THE URETERAL ORIFICES (UOS). DUE TO THE PERSISTENT ISSUES, THE PROCEDURE WAS ABORTED WHILE THE PATIENT WAS UNDER ANESTHESIA. THERE WERE NO ADVERSE HEALTH CONSEQUENCES FOR THE PATIENT AS A RESULT OF THIS EVENT.

Additional Manufacturer Narrative

ROOT CAUSE OF THE REPORTED EVENT HAS NOT YET BEEN ESTABLISHED. INVESTIGATION BY MANUFACTURER IS CURRENTLY IN-PROCESS.SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.