UNIDENTIFIED
Received Jul 16, 2026 · Event occurred Mar 23, 2026
Report 1000513161-2026-00017 · MDR key 25840605
Device
Generic name
Insufflator, Automatic Carbon-dioxide For Endoscope
Product problems
- Explosion
Patient
NA · Female
- Acoustic Trauma
- Partial Hearing Loss
Narrative
Description of Event or Problem
ON OR AROUND 06/25/2026, FUJIFILM HEALTHCARE AMERICAS CORPORATION BECAME AWARE OF AN INCIDENT INVOLVING A FUJIFILM DEVICE DURING A DEMO. ON OR ABOUT 03/23/2026, THE HCUS CLINICAL SPECIALIST WAS AT (B)(6) HOSPITAL - PLANO (THE "HOSPITAL") TO DEMONSTRATE HOW TO USE NEW FUJIFILM EQUIPMENT TO THE HOSPITAL STAFF. THE DEMONSTRATION USED A FUJIFILM ENDOSCOPE IN CONNECTION WITH AN ENDOSCOPIC ULTRASOUND AND ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPLIY PROCEDURE ("EUS/ERCP"). AN EUS/ERCP INVOLVES A STERILE WATER BOTTLE CONNECTED TO A CARBON-DIOXIDE INSUFFLATOR. THE INSUFFLATOR DELIVERS A CONTINUOUS FLOW OF CARBON-DIOXIDE GAS INTO THE WATER BOTTLE TO FACILITATE LENS CLEANING AND AUXILIARY IRRIGATION. DURING THE DEMONSTRATION, THE HCUS CLINICAL SPECIALIST CLAMPED THE STERILE WATER BOTTLE TUBING, PREVENTING THE BOTTLE FROM VENTING EXCESS PRESSURE. THE TRAPPED PRESSURIZED GAS CAUSED THE BOTTLE TO BURST (THE "EXPLOSION"). A HOSPITAL STAFF IN ATTENDANCE FOR THE DEMONSTRATION WAS WEARING HEARING AIDS AT THE TIME OF THE EXPLOSION. THE SOUND OF THE EXPLOSION, WHICH WAS MAGNIFIED BY TBE HEARING AIDS, SEVERELY DAMAGED THE HOSPITAL STAFF'S HEARING. FUJIFILM HEALTHCARE AMERICAS CORPORATION IS SUBMITTING THE REPORT DUE TO THE ALLEGED SEVERE DAMAGE TO THE HOSPITAL STAFF'S HEARING. REF: FUJIFILM HEALTHCARE AMERICAS CORPORATION COMPLAINT # (B)(4).