inforMED
InjuryDRX

SKINTACT

Received Jul 16, 2026

Report 8020045-2026-00021 · MDR key 25835655

Device

Generic name

Ecg Electrode

Manufacturer

Leonhard Lang Gmbh

Model number

TVO01

Lot number

250617-0243

Product problems

  • Insufficient Device Problem Information

Patient

80 YR · Male

  • Localized Skin Lesion
  • Skin Inflammation/ Irritation

Narrative

Description of Event or Problem

ON (B)(6) 2026, WE HAVE BEEN INFORMED ABOUT AN INCIDENT WITH ECG ELECTRODES. MONITORING ECG ELECTRODES (MODEL TVO01) HAD BEEN USED AT HERZPRAXIS (B)(6). THE INITIAL REPORTER INFORMED LEONHARD LANG ABOUT A PATIENT INCIDENT. IT WAS STATED: "A CARDIOLOGY PRACTICE HAS COMPLAINED THAT PATIENTS FREQUENTLY REPORT ITCHING AND REDNESS FROM THE ELECTRODES, WHICH CAN REMAIN VISIBLE FOR DAYS. TWO PATIENTS EVEN DEVELOPED BLISTERS ON THEIR SKIN." ON (B)(6) 2026 WE HAVE RECEIVED A FILLED IN QUESTIONAIRE. A 48-HOUR ECG MONITORING WAS PERFORMED. THE PATIENT WAS DESCRIBED AS SLIM AND THE PATIENT SKIN WAS DESCRIBED AS NORMAL, "RATHER DRY, BODY HAIR? NO TATTOOS". A MORTARA H12+ ECG DEVICE AND MEDILAN CABLES HAVE BEEN USED. 7 ECG ELECTRODE HAVE BEEN APPLIED ON THE PATIENT'S CHEST. THE SKIN PREPARATION WAS DESCRIBED AS CLEANED WITH "ALCOHOL ISOPROPYLICUS PROPAN-2-OL", SHAVEN? DISINFECTED WITH "ALCOHOL ISOPROPYLICUS PROPAN-2-OL", NO OINTMENT HAS BEEN USED BUT THE SKIN WAS DRIED. IT WAS ALSO REPORTED THAT " THE ADHESIVE STRENGTH IS GOOD, BUT SKIN REACTION IS A PROBLEM." THE ELECTRODES ADHERED WELL AND WERE NOT REPOSITIONED. ON (B)(6) 2026 WE HAVE RECEIVED FURTHER INFORMATION AND TWO PICTURES SHOWING THE PATIENT INJURY AT THE COLLARBONE. THE PICTURES ARE SHOWING A REDDENED LONGISH AREA OF THE PATIENT SKIN. A PRECISE ESTIMATION OF THE SIZE IS NOT POSSIBLE DUE TO MISSING SIZE COMPARISON. IT WAS ALSO STATED THAT "THE CAUSE OF THIS PATIENT'S SEVERE REACTION MAY ALSO BE PARTLY DUE TO THE APPLICATION OF THE ADHESIVE. MEDILAN INSTRUCTED US TO ALWAYS PERFORM A GENTLE EXFOLIATION BEFORE APPLYING THE ELECTRODES. TO BE HONEST, THE EXFOLIATION WAS PROBABLY A BIT TOO VIGOROUS IN THIS PATIENT'S CASE, WHICH CERTAINLY CONTRIBUTED TO THE SEVERE REACTION. HOWEVER, SINCE WE HAVE BEEN RECEIVING INCREASING REPORTS OF ALLERGIC REACTIONS FROM OTHER PATIENTS SINCE SPRING, AND WE WERE ALSO APPROACHED BY A PARTNER MEDICAL PRACTICE THAT USES THE SAME ELECTRODES, I DECIDED TO REPORT THIS TO YOU." NO INFORMATION WAS PROVIDED IF AND HOW THE INJURY HAS TO BE TREATED AFTERWARDS. WE HAVE REQUESSTED FURTHER INFORMATION AND CUSTOMER SAMPLES AND HAVE CURRENTLY BEEN INFORMED THAT "THE PRACTICE IS CLOSED FOR 3 WEEKS DUE TO VACATION". THEREFORE WE WILL PROVIDE FURTHER INFORMATION AND ANY CONCLUSION IN A FOLLOW UP REPORT ONCE THE PRACTICE HAS OPEN AGAIN AFTER VACATION.

Additional Manufacturer Narrative

RETAINED SAMPLES OF THE CONCERNED LOT OF MODEL TVO01 HAVE BEEN INSPECTED VISUALLY. ALL INSPECTED SAMPLES WERE FOUND TO PERFORM WITHIN LIMITS. WE ALSO HAVE REVIEWED THE PRODUCTION HISTORY RECORDS FOR THE CONCERNED LOT NUMBER. MECHANICAL TESTS WERE PERFORMED ON 3 RETAINED SAMPLES. ALL TESTED SAMPLES WERE FOUND TO PERFORM WITHIN LIMITS. ADDITIONALLY A SKIN ADHESION TEST ON A TEST PERSON WAS PERFORMED FOR ABOUT 1 HOUR STICKING THE ECG TEST ELECTRODES OF THE CONCERNED LOT NUMBER TO THE SKIN. NO FAULTS COULD BE DETECTED. WE HAVE REQUESTED FOR CUSTOMER SAMPLES AND IF THE PATIENT INJURY HAS TO BE TREATED AFTERWARDS BUT NONE HAVE BEEN RECEIVED SO FAR AS IT WAS STATED THAT THE PRACTICE IS CLOSED FOR 3 WEEKS DUE TO VACATION. THEREFORE WE WILL PROVIDE FURTHER INFORMATION AND ANY CONCLUSION IN A FOLLOW UP REPORT.