MalfunctionFDF

EVIS LUCERA COLONOVIDEOSCOPE

Received Jul 15, 2026 · Event occurred Apr 16, 2026

Report 9610595-2026-56985 · MDR key 25829772

Device

Generic name

Colonovideoscope

Model number

PCF-Q260AZI

Product problems

  • Failure to Clean Adequately

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION. THE MOST PROBABLE CAUSE OF THE COMPLAINT IS CAUSE NOT ESTABLISHED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem

THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE TEAM INDICATES THAT CRYSTALLIZATION IN BOTH THE CONTROL UNIT AND THE OPTICAL WAVEGUIDE ROD AND WATER SCALE BUILDUP UNDER THE OPTICAL WAVEGUIDE LENS WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.