DOUBLE MOBILITY LINER
Received Jul 15, 2026 · Event occurred Jul 1, 2026
Report 3005180920-2026-00698 · MDR key 25822786
Device
Generic name
Double Mobility Hc Liner Ø28/dmf
Manufacturer
Medacta International SaModel number
01.26.2852MHCLot number
2601010
Product problems
- Device Appears to Trigger Rejection
Patient
58 YR · Female
- Medical device site infection
Narrative
Additional Manufacturer Narrative
BATCH REVIEW PERFORMED ON 06 JUL 2026 LINER: MPACT DM 01.26.2852MHC DOUBLE MOBILITY HC LINER D 28/DMF (K092265) LOT 2601010: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10-MAR-2026. EXPIRATION DATE: 2031-FEB-10. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. BALL HEADS: MECTACER 01.29.230H MECTACER HEAD BIOLOX OPTION 12/14 D 28 (K131518) LOT. 2519245: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 02-SEPT-2025. EXPIRATION DATE: 2030-JUL-17. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. BALL HEADS: MECTACER 01.29.243A MECTACER SLEEVE 12/14 SIZE XL (K131518) LOT. 2515762: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 30-JUN-2025. EXPIRATION DATE: 2030-JUN-11. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ROOT CAUSE: INFECTION IS A KNOWN POSSIBLE COMPLICATION OF ANY SURGERY. ALTHOUGH NO ROOT CAUSE CAN BE ESTABLISHED, THERE IS NO INDICATION THAT ANY ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DOCUMENT REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED CAUSE.
Description of Event or Problem
AT ABOUT ONE MONTH FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE DOUBLE MOBILITY HC LINER D 28/DMF, 28MM BIOLOX OPTION HEAD, AND BIOLOX OPTION SLEEVE XL TO A SAME SIZE LINER, HEAD, AND SLEEVE. THE SURGERY WAS COMPLETED SUCCESSFULLY.