inforMED
MalfunctionLNH

INGENIA

Received Jul 15, 2026 · Event occurred Jul 3, 2026

Report 3042177665-2026-000110 · MDR key 25822618

Device

Generic name

Full-body Mri System, Superconducting Magnet

Catalog number

781260

Product problems

  • Detachment of Device or Device Component

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL HAS COME LOOSE THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS DAMAGED IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

Additional Manufacturer Narrative

THE QUADRATURE BODY COIL (QBC) SEAL HAS COME LOOSE AND WAS CORRECTED BY IMPLEMENTING AN FCO.THE FAILURE OF THE QBC SEAL TO REMAIN IN PLACE IS CONSIDERED A MALFUNCTION.REMEDIAL ACTION:A FIELD ACTION WAS INITIATED UNDER 2023-PD-MR-013.A FIELD SAFETY NOTIFICATION WAS SENT TO THE CUSTOMERS INFORMING THEM ABOUT THIS POTENTIAL ISSUE.THE FIELD CORRECTION HAS SINCE BEEN MADE AVAILABLE TO THE FIELD IN DEC 2024.

Remedial action

  • Modification/Adjustment