inforMED
MalfunctionHXX

INVICTUS SPINAL FIXATION SYSTEM

Received Jul 14, 2026 · Event occurred Jun 24, 2026

Report 2027467-2026-00153 · MDR key 25813997

Device

Generic name

Screwdriver

Model number

17950-225

Catalog number

17950-225

Lot number

EM49771

Product problems

  • Break
  • Material Separation

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

VISUAL INSPECTION CONFIRMS THE EVENT. THE DISTAL TIP HAS SHEARED OFF AT THE BASE OF THE HEXALOBE. THIS FAILURE IS LIKELY DUE TO THE APPLIED TORSIONAL FORCE BEING GREATER THAN THE YIELD STRENGTH OF THE DEVICE. REVIEW OF DEVICE HISTORY RECORDS SHOWED NO MANUFACTURING OR PROCESSING RELATED IRREGULARITIES. LABELING REVIEW: WARNINGS/CAUTIONS/PRECAUTIONS: RISKS IDENTIFIED WITH THE USE OF THESE DEVICES, WHICH MAY REQUIRE ADDITIONAL SURGERY, INCLUDE DEVICE COMPONENT FAILURE, LOSS OF FIXATION/STABILIZATION, NON-UNION, VERTEBRAL FRACTURE, NEUROLOGICAL INJURY, VASCULAR OR VISCERAL INJURY. POSSIBLE ADVERSE EFFECTS: INITIAL OR DELAYED LOOSENING, DISASSEMBLY, BENDING, DISLOCATION, AND/OR BREAKAGE OF DEVICE COMPONENTS.

Description of Event or Problem

IT WAS REPORTED THAT THE DRIVER BROKE DURING CLEANING AFTER A CASE.