inforMED
InjuryJWH

GMK PRIMARY TOTAL KNEE SYSTEM

Received Jul 14, 2026 · Event occurred Jul 2, 2026

Report 3005180920-2026-00725 · MDR key 25813729

Device

Generic name

Tibial Tray Fix Cemented S.3l

Model number

02.07.1203L

Lot number

2010232

Product problems

  • Loosening of Implant Not Related to Bone-Ingrowth

Patient

NA · Male

  • Aseptic loosening

Narrative

Description of Event or Problem

AT ABOUT 5 YEARS FROM THE PRIMARY, THE PATIENT CAME IN PRESENTING ASEPTIC LOOSENING OF THE TIBIA AND THE CAUSE IS UNKNOWN. THERE WAS NO INDICATION OF TRAUMA OR A FALL. THE SURGEON REVISED THE GMK-SPHERE TIBIAL TRAY, INSERT, AND FEMORAL COMPONENT TO GMK-HINGE IMPLANTS. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Additional Manufacturer Narrative

BATCH REVIEW PERFORMED ON 06 JULY 2026 GMK-SPHERE 02.07.1203L TIBIAL TRAY FIX CEMENTED S.3L (K090988) LOT 2010232: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 02-MAR-2021. EXPIRATION DATE: 17-FEB-2026. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ROOT CAUSE:ASEPTIC LOOSENING IS POSSIBLE LITERATURE DESCRIBED ADVERSE EVENT AFTER PRIMARY JOINT ARTHROPLASTIES AND CAUSES ARE OFTEN UNKNOWN. THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AND THE INVESTIGATION DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED ISSUE OR SYSTEMIC DEFICIENCY.