inforMED
InjuryJCW

AMS 700? CX TENACIO? PUMP

Received Jul 13, 2026 · Event occurred Jun 17, 2026

Report 2124215-2026-37518 · MDR key 25806668

Device

Generic name

Prosthesis, Penis, Inflatable

Model number

72404434

Catalog number

72404434

Lot number

1100784774

Product problems

  • Material Puncture/Hole
  • Migration

Patient

64 YR · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

MODEL NUMBER/CATALOG NUMBER: 720185-01. SERIAL NUMBER: N/A. BATCH/LOT NUMBER: 1100812152. MODEL/CATALOG DESCRIPTION: RESERVOIR FLAT IZ 100 ML. UNIQUE IDENTIFIER (UDI) #: (B)(4).

Description of Event or Problem

IT WAS REPORTED THAT THE DISTAL END OF THE LEFT CYLINDER OF AN INFLATABLE PENILE PROSTHESIS (IPP) HAD MIGRATED OUT OF THE CORPOROTOMY TO THE SCROTUM. UPON EXPLANTATION, IT WAS FOUND THAT THE OUTER SILICONE LAYER HAD RUPTURED. AS A RESULT, THE ENTIRE DEVICE WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.