DRILL, FOR FIBERTAK DR
Received Jul 13, 2026 · Event occurred May 26, 2026
Report 1220246-2026-03749 · MDR key 25798627
Device
Generic name
Drill Bit
Manufacturer
Arthrex, Inc.Model number
DRILL, FOR FIBERTAK DRCatalog number
AR-3657
Lot number
10491659
Product problems
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE REPORTED EVENT WAS CONFIRMED AS THE VISUAL EVALUATION OF THE RECEIVED AR-3657 WITH BATCH 10491659 NOTED THAT THE CUTTING-EDGE TIP IS BROKEN OFF (SEE EVALUATION PICTURES 3 + 4). NO BROKEN PIECES WERE RETURNED FOR INVESTIGATION. A FUNCTIONAL TEST WAS NOT PERFORMED DUE TO DAMAGE TO THE DEVICE. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO USER ERROR OF THE DEVICE DUE TO APPLYING EXCESSIVE MECHANICAL FORCES.
Description of Event or Problem
IT WAS REPORTED THAT DURING THE ROTATOR CUFF SPEEDFIX SURGERY, THE TIP OF THE DRILL BROKE OFF. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.