TEMPO AQUA
Received Jul 13, 2026 · Event occurred Jun 22, 2026
Report 9616099-2026-00966 · MDR key 25796686
Device
Generic name
Catheter, Intravascular, Diagnostic
Manufacturer
Cordis Us. CorpCatalog number
452531H0
Lot number
18278852
Product problems
- Detachment of Device or Device Component
Patient
82 YR · Male
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS DEVICE IS AVAILABLE FOR ANALYSIS, BUT THE ENGINEERING REPORT IS NOT YET AVAILABLE. HOWEVER, IT WILL BE SUBMITTED WITHIN 30 DAYS UPON RECEIPT. ADDITIONAL INFORMATION IS PENDING AND WILL BE SUBMITTED WITHIN 30 DAYS UPON RECEIPT.
Description of Event or Problem
AS REPORTED, THERE WAS DIFFICULTY ADVANCING A 5F 100CM SIDEWINDER SIMMONS II (SIM2) TEMPO AQUA DIAGNOSTIC CATHETER TOWARD THE LESION AND THE DEVICE FRACTURED IN THE AORTIC ARCH. THE FRACTURE CAUSED THE DEVICE TO SEPARATE INTO MULTIPLE PIECES, AND THE SEPARATED CATHETER SEGMENT WAS REMOVED THROUGH EMERGENCY MINIMALLY INVASIVE INTERVENTION. THE DEVICE WAS STORED AND PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), WITH NO ATTEMPT TO SHAPE IT DURING PREPARATION. THE FACILITY DID NOT USE ULTRAVIOLET (UV) LIGHT OR OZONE FOR ROOM STERILIZATION. THE INTENDED PROCEDURE WAS CAROTID ARTERY STENTING (CAS) VIA FEMORAL ARTERY ACCESS. WITHIN THE AORTIC ARCH, THERE WAS NO CALCIFICATION, TORTUOSITY, STENOSIS, OR ACUTE ANGLE. THE DEVICE WAS IN AN ACUTE BEND DURING THE PROCEDURE, WITH NO INDICATION OF CATHETER KINKING PRIOR TO FRACTURE. THE DEVICE WILL BE RETURNED FOR EVALUATION.