Device

Generic name

Enzymatic, Carbon-dioxide

Manufacturer

Abbott Gmbh

Catalog number

07P72-20

Lot number

69531UQ01

Product problems

  • Low Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Description of Event or Problem

THE CUSTOMER OBSERVED FALSELY DECREASED CARBON DIOXIDE (CO2) RESULTS GENERATED FROM ALINITY C PROCESSING MODULE FOR MULTIPLE PATIENTS. THE RESULTS PROVIDED WERE: INITIAL=14.0 MMOL/L /REPEATED=23.0 MMOL/L ON (B)(6) 2026 SID (B)(6): 60YRS OLD MALE: INITIAL=17 MMOL/L /REPEATED=22 MMOL/L LABORATORY REFERENCE RANGE FOR CO2=22 TO 30 MMOL/L THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.