inforMED
MalfunctionCAD

TEC

Received Jul 13, 2026 · Event occurred Jun 15, 2026

Report 2112667-2026-04426 · MDR key 25795823

Device

Generic name

Vaporizer

Manufacturer

Datex-ohmeda, Inc.

Model number

850 SEV

Catalog number

NA

Lot number

NA

Product problems

  • Insufficient Flow or Under Infusion

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE ACTUATOR ASSEMBLY WAS CALIBRATED TO RESOLVE THE ISSUE. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS MADISON - 3030 OHMEDA DR, USA MADISON, WI 53718.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT AGENT DELIVERY LESS THAT -20% OF SETTING. THERE WAS NO PATIENT INVOLVEMENT.