CONCEAL
Received Jul 10, 2026 · Event occurred Jan 1, 2025
Report 2124215-2026-37224 · MDR key 25786554
Device
Generic name
Prosthesis, Penis, Inflatable
Manufacturer
Boston Scientific CorporationModel number
720185-01Catalog number
720185-01
Lot number
1100496358
Product problems
- Migration
Patient
63 YR · Male
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
MODEL NUMBER/CATALOG NUMBER: 72404267. SERIAL NUMBER: N/A. BATCH/LOT NUMBER: 1100526411. MODEL/CATALOG DESCRIPTION: CXR PRECONNECT MS 14CM IP IZ. UNIQUE IDENTIFIER (UDI) #: (B)(4). THE DEVICE IS NOT AVAILABLE FOR ANALYSIS; THEREFORE, NO PHYSICAL ANALYSIS OF THE PRODUCT COULD BE PERFORMED. THE REPORTED DEVICE PERFORMANCE ALLEGATION CANNOT BE CONFIRMED. MIGRATION IS ACKNOWLEDGED WITHIN THE INSTRUCTIONS FOR USE (IFU) AND IS NOT RELATED TO OFF-LABEL USE OR FAILURE TO FOLLOW INSTRUCTIONS.
Description of Event or Problem
IT WAS REPORTED THAT A PATIENT WITH AN INFLATABLE PENILE PROSTHESIS (IPP) PRESENTED WITH A BULGE IN THE LOWER ABDOMEN UNDER THE SKIN. A REVISION SURGERY WAS PERFORMED, DURING WHICH THE PHYSICIAN CONFIRMED THAT THE RESERVOIR HAD MIGRATED FROM ITS ORIGINAL LOCATION IN THE SPACE OF RETZIUS. THE RESERVOIR WAS EXPLANTED AND REPLACED WITH A NEW RESERVOIR POSITIONED UNDER THE LEFT RECTUS MUSCLE. THERE WERE NO PATIENT COMPLICATIONS.