SHOCKWAVE IVL SYSTEM WITH THE SHOCKWAVE M5+ PERIPHERAL IVL CATHETER
Received Jul 10, 2026 · Event occurred Jun 22, 2026
Report 3015053858-2026-00075 · MDR key 25785239
Device
Generic name
Shockwave Ivl System With The Shockwave M5+ Peripheral Ivl Catheter
Manufacturer
Shockwave Medical, Inc.Model number
M5PIVL6060Catalog number
M5PIVL6060
Lot number
82362390
Product problems
- Detachment of Device or Device Component
Patient
Not reported
- Foreign Body In Patient
Narrative
Additional Manufacturer Narrative
THE DEVICE REFERENCED IN THIS REPORT WAS NOT RETURNED TO SHOCKWAVE MEDICAL, INC. THEREFORE, A PHYSICAL EXAMINATION CANNOT BE PERFORMED. THE CAUSE OF THE BALLOON DETACHMENT COULD NOT BE DEFINITIVELY DETERMINED BASED ON THE INFORMATION AVAILABLE. SHOCKWAVE MEDICAL HAS CONTROLS IN PLACE TO ENSURE DEVICES ARE BUILT TO APPROVED PROCEDURES AND MEET LOT RELEASE ACCEPTANCE CRITERIA PRIOR TO BEING DISTRIBUTED. A REVIEW OF THE MANUFACTURING AND TEST DOCUMENTATION FOR SUBJECT LOT DOES NOT REVEAL ANY ISSUES WITH THE MANUFACTURING OF THE DEVICE. THE DEVICE PASSED ALL OF SHOCKWAVE MEDICAL, INC. ACCEPTANCE CRITERIA PRIOR TO SHIPPING.
Description of Event or Problem
A SHOCKWAVE M5+ PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED TO TREAT A LESION IN THE SUPERFICIAL FEMORAL ARTERY/POPLITEAL ARTERY. FOLLOWING SUCCESSFUL IVL TREATMENT AND DURING DEVICE REMOVAL AT THE PROXIMAL ANTERIOR TIBIAL (AT) ARTERY, A PORTION OF THE BALLOON DETACHED FROM THE CATHETER. APPROXIMATELY 2 CM OF THE BALLOON SEPARATED AND COULD NOT BE RETRIEVED, AND WAS SUBSEQUENTLY STENTED AGAINST THE VESSEL WALL. THE BALLOON WAS REPORTED TO BE FULLY DEFLATED PRIOR TO REMOVAL, WITH NO BALLOON RUPTURE OR PRESSURE LOSS OBSERVED. NO EXCESSIVE FORCE WAS REPORTEDLY USED DURING DEVICE MANIPULATION OR REMOVAL. NO PATIENT HARM WAS REPORTED.