APTIMA MULTITEST SPECIMEN COLLECTION KIT
Received Jul 10, 2026 · Event occurred Jun 15, 2026
Report 2024800-2026-00042 · MDR key 25782390
Device
Generic name
Device, Specimen Collection
Manufacturer
Hologic, Inc.Catalog number
ASY-08848
Lot number
932841
Product problems
- Material Separation
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
ON (B)(6) 2026, A CUSTOMER FROM NETHERLANDS REPORTED TO HOLOGIC THAT THE TIP OF A SWAB, INCLUDED IN THE APTIMA MULTITEST SPECIMEN COLLECTION KIT (LOT NUMBER: 932841) DETACHED FROM THE SWAB SHAFT AND REMAINED INSIDE THE PATIENT DURING RECTAL SPECIMEN COLLECTION. THE CUSTOMER STATED THAT THE PATIENT REMOVED THE SWAB TIP THEMSELVES. NO FURTHER INFORMATION ON THE PATIENT¿S STATUS WAS PROVIDED TO HOLOGIC. HOLOGIC HAS NOT LEARNED OF ANY ADVERSE OUTCOMES RELATED TO THIS EVENT.
Additional Manufacturer Narrative
A REVIEW OF THE PART HISTORY FILE ASSOCIATED WITH THE MULTITEST SWABS INCLUDED IN THE APTIMA MULTITEST SWAB KIT (LN: 932841) CONFIRMED THAT THE SWABS SUCCESSFULLY PASSED INSPECTION SPECIFICATIONS. HOLOGIC HAS NOT BEEN INFORMED OF ANY ADVERSE OUTCOMES RELATED TO THIS ISSUE. A SUPPLIER CORRECTIVE ACTION REQUEST (SCAR) HAS BEEN ISSUED TO THE SUPPLIER OF THE SWABS (B)(4). H3 OTHER TEXT: OTHER.