MalfunctionGCM

VIDEO-OPTIK "ENDOEYE 3D", 30°

Received Jul 10, 2026

Report 9610773-2026-04857 · MDR key 25781321

Device

Generic name

Rigid Video Laparoscope

Model number

WA50082A

Product problems

  • Corroded

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE CAUSE OF THE CORROSION ON THE R-UNIT (CHARGED COUPLED DEVICE) TUBE WAS DUE TO COMPONENT FAILURE OF THE INSERTION TUBE. DATE OF EVENT IS UNKNOWN. ADDITIONAL INFORMATION WILL BE PROVIDED IF AVAILABLE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem

THE CUSTOMER RETURNED THE RIGID VIDEO LAPAROSCOPE, WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, CORROSION ON THE R-UNIT (CHARGED COUPLED DEVICE) TUBE WAS OBSERVED. THERE WAS NO PATIENT INVOLVEMENT.