MECTACER BIOLOX DELTA FEMORAL BALL HEAD
Received Jul 10, 2026 · Event occurred Jun 23, 2026
Report 3005180920-2026-00690 · MDR key 25778108
Device
Generic name
Mectacer Head Biolox Delta Dia.36 12/14-s
Manufacturer
Medacta International SaModel number
01.29.208Lot number
2532288
Product problems
- Device Dislodged or Dislocated
Patient
NA · Female
- Joint Dislocation
Narrative
Description of Event or Problem
REVISION SURGERY DUE TO JOINT LUXATION AFTER ABOUT 3 MONTHS FROM PRIMARY. THE SURGEON REVISED THE 36 MM BIOLOX DELTA HEAD S TO A 28 MM BIOLOX OPTION HEAD WITH SLEEVE L AND REVISED THE FLAT PE HC LINER D 36/E TO A DOUBLE MOBILITY HC LINER D 28/DMC ALONG WITH A DM CONVERTER.
Additional Manufacturer Narrative
BATCH REVIEW PERFORMED ON 09 JULY 2026: MECTACER 01.29.208 MECTACER HEAD BIOLOX DELTA DIA.36 12/14-S (K112115) LOT. 2532288: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 24-DEC-2025. EXPIRATION DATE: 2030-DEC-11. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. MPACT 01.32.3644HCT FLAT PE HC LINER D 36/E (K103721) LOT. 2528030: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 07-JAN-2026. EXPIRATION DATE: 2030-NOV-25. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ROOT CAUSE: BASED ON THE INFORMATION AVAILABLE NO DEFINITIVE ROOT CAUSE CAN BE ESTABLISHED, WHILE THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DEVICE HISTORY RECORD REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED ISSUE.