inforMED
InjuryDZL

LOCKSCREW,AXS,ST,2.0X10MM,5P

Received Jul 10, 2026 · Event occurred Jun 15, 2026

Report 0008010177-2026-00096 · MDR key 25776259

Add to watch list
Product code (DZL)
Model (56-20510)

Device

Generic name

Screw, Fixation, Intraosseous

Catalog number

56-20510

Lot number

UNKNOWN

Product problems

  • Fracture

Patient

NA · Unknown

  • Insufficient Information

Narrative

Description of Event or Problem

IT WAS REPORTED THAT APPROXIMATELY SIX WEEKS POSTOPERATIVELY, THE PLATE BECAME LOOSE DUE TO SCREW BACKOUT. DURING REVISION SURGERY, A SHEARED SCREW HEAD WAS IDENTIFIED, REQUIRING REPLACEMENT OF THE PLATE AND SCREWS. THE SURGEON ATTRIBUTED THE FAILURE TO PATIENT BONE QUALITY RATHER THAN A DEVICE RELATED ISSUE.