IMAGE1 S D3-LINK
Received Jul 9, 2026 · Event occurred Jun 22, 2026
Report 9610617-2026-01585 · MDR key 25766541
Device
Generic name
Image1 S D3-link
Manufacturer
Karl Storz Se & Co. KgModel number
26605BACatalog number
26605BA
Product problems
- Erratic or Intermittent Display
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
ACCORDING TO THE INFORMATION RECEIVED, ON (b)(6) 2026, THE PATIENT UNDERWENT LAPAROSCOPIC CHOLECYSTECTOMY FOR GALLSTONES IN THE OPERATING ROOM. THE IMAGE FLICKERED AND AN IDENTIFICATION ERROR ALARM APPEARED DURING THE OPERATION, CAUSING A BRIEF INTERRUPTION OF THE SURGERY. THE ISSUE WAS ATTRIBUTED TO A SIGNAL ACQUISITION AND TRANSMISSION FAILURE OF THE CONNECTING CABLE, WHICH RESULTED IN SCREEN FLICKERING AND ERROR ALERTS. THE SURGERY WAS COMPLETED AFTER EMERGENCY HANDLING. THE DEVICE HAS BEEN TAKEN OUT OF SERVICE FOR MAINTENANCE AND WILL BE EVALUATED FOR USE AFTER THE CONNECTING CABLE IS REPLACED. THIS INCIDENT INVOLVED A BRIEF SURGICAL INTERRUPTION WITH SIGNIFICANT RISKS, SO IT IS CLASSIFIED AS SEVERE. ADDITIONAL INFORMATION RECEIVED: THE ENDOSCOPE SCREEN FLICKERED MOMENTARILY DURING SURGERY, THEN RETURNED TO NORMAL IMMEDIATELY. THE PROCEDURE WAS COMPLETED WITH THIS ENDOSCOPE WITHOUT SIGNIFICANT DELAY, ONLY LASTING A FEW SECONDS.
Additional Manufacturer Narrative
UNDER USER'S DISCRETION THE DEVICE WILL NOT BE RETURNED TO THE MANUFACTURER FOR EVALUATION. THE INVESTIGATION WILL BE PERFORMED BY A DESIGNATED KARL STORZ EMPLOYEE, INVESTIGATION RESULTS ARE PENDING.THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID: (b)(4).