inforMED
InjuryPPN

SHOCKWAVE IVL SYSTEM WITH THE SHOCKWAVE M5+ PERIPHERAL IVL CATHETER

Received Jul 9, 2026 · Event occurred Jun 24, 2026

Report 3015053858-2026-00074 · MDR key 25765291

Device

Generic name

Shockwave Ivl System With The Shockwave M5+ Peripheral Ivl Catheter

Model number

M5PIVL6060

Catalog number

M5PIVL6060

Lot number

82364266

Product problems

  • Detachment of Device or Device Component
  • Material Split, Cut or Torn

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT A SHOCKWAVE M5+ PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED DURING A PROCEDURE TO TREAT A LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA)/POPLITEAL ARTERY. THE IVL BALLOON RUPTURED FOLLOWED BY PARTIAL DETACHMENT WHILE IT WAS BEING PULLED OUT OF A 5F SHORT SHEATH; THE M5+ IVL IFU SPECIFIES THE USE OF A 6F SHEATH. THE PATIENT WAS TAKEN TO THE OPERATING ROOM (OR) FOR A SURGICAL CUT-DOWN PROCEDURE TO REMOVE THE DETACHED PORTION OF THE IVL BALLOON. NO PATIENT HARM WAS REPORTED. THE PATIENT WAS DISCHARGED, AND NO OTHER ISSUES REPORTED.

Additional Manufacturer Narrative

THE DEVICE REFERENCED IN THIS REPORT HAS NOT YET BEEN RECEIVED AT SHOCKWAVE MEDICAL, INC. THEREFORE, A PHYSICAL EXAMINATION HAS NOT BEEN PERFORMED. EVALUATION OF THE SUBJECT DEVICE IS ANTICIPATED BUT HAS NOT YET BEGUN. AFTER THE DEVICE IS RECEIVED AND INVESTIGATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.