GMK SPHERE TOTAL KNEE SYSTEM
Received Jul 9, 2026 · Event occurred Jun 23, 2026
Report 3005180920-2026-00674 · MDR key 25763496
Device
Generic name
Gmk-sphere Tibial Insert E-cross - Flex 5r - 10mm
Manufacturer
Medacta International SaModel number
02.12.E0510FRLot number
2521356
Product problems
- Device Appears to Trigger Rejection
Patient
64 YR · Male
- Medical device site infection
Narrative
Additional Manufacturer Narrative
BATCH REVIEW PERFORMED ON 07 JULY 2026 LOT. 2521356: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 30-SEP-2025. EXPIRATION DATE: 2030-SEP-16. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH ONE SIMILAR REPORTED EVENTS DURING THE PERIOD OF REVIEW. ROOT CAUSE: INFECTION IS A KNOWN POSSIBLE COMPLICATION OF ANY SURGERY. ALTHOUGH NO ROOT CAUSE CAN BE ESTABLISHED, THERE IS NO INDICATION THAT ANY ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DOCUMENT REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED CAUSE.
Description of Event or Problem
REVISON SURGERY DUE TO INFECTION AFTER ABOUT 1 MONTH FROM PRIMARY. THE SURGEON PERFORMED A WASHOUT AND REVISED THE INSERT TO A SAME SIZE INSERT.