VERSAFITCUP CC TRIO ACETABULAR SYSTEM
Received Jul 9, 2026 · Event occurred Jun 22, 2026
Report 3005180920-2026-00685 · MDR key 25763494
Device
Generic name
Versafitcup Metalback Cc Trio Ø48 Mm
Manufacturer
Medacta International SaModel number
01.26.45.0048Lot number
2006316
Product problems
- Loosening of Implant Not Related to Bone-Ingrowth
Patient
76 YR · Female
- Aseptic loosening
Narrative
Description of Event or Problem
REVISION SURGERY DUE TO A LOOSE CUP AFTER ABOUT 5 YEARS AND 3 MONTHS. THE SURGEON REVISED THE VERSAFITCUP METALBACK CC TRIO D 48 MM TO A MPACT 3D METAL SHELL TWO HOLE D 52MM, REVISED THE FLAT PE HC LINER D 32/C TO A DOUBLE MOBILITY HC LINER D 28/DMC WITH DM CONVERTER, AND REVISED THE 32MM BIOLOX DELTA HEAD M TO A 28MM BIOLOX OPTION HEAD WITH SLEEVE. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Additional Manufacturer Narrative
BATCH REVIEW PERFORMED ON 09 JULY 2026 LOT 2006316: (B)(6) ITEMS MANUFACTURED AND RELEASED ON 02-OCT-2020. EXPIRATION DATE: 23-SEP-2025. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(6) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ROOT CAUSE: ASEPTIC LOOSENING IS A POSSIBLE LITERATURE-DESCRIBED ADVERSE EVENT AFTER PRIMARY HIP ARTHROPLASTIES AND CAUSES ARE OFTEN UNKNOWN. THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT.