AMS 700 CX MS PUMP
Received Jul 8, 2026 · Event occurred Jun 1, 2026
Report 2124215-2026-36635 · MDR key 25751088
Device
Generic name
Prosthesis, Penis, Inflatable
Manufacturer
Boston Scientific CorporationModel number
72404233-12Catalog number
72404233-12
Lot number
1100505748
Product problems
- Deflation Problem
- Mechanical Problem
- Migration
Patient
Not reported
- Perforation
Narrative
Additional Manufacturer Narrative
MODEL NUMBER/CATALOG NUMBER: 72404156. SERIAL NUMBER: N/A. BATCH/LOT NUMBER: 1100754237. MODEL/CATALOG DESCRIPTION: RESERVOIR 100 ML PC IZ. UNIQUE IDENTIFIER (UDI) #: (B)(4). THE DEVICE IS NOT AVAILABLE FOR ANALYSIS; THEREFORE, NO PHYSICAL ANALYSIS OF THE PRODUCT COULD BE PERFORMED. THE REPORTED DEVICE PERFORMANCE ALLEGATION CANNOT BE CONFIRMED. DEFLATION ISSUE, MECHANICAL ISSUE, MIGRATION AND PERFORATION ARE ACKNOWLEDGED WITHIN THE INSTRUCTIONS FOR USE (IFU) AND ARE NOT RELATED TO OFF-LABEL USE OR FAILURE TO FOLLOW INSTRUCTIONS.
Description of Event or Problem
IT WAS REPORTED THAT A PATIENT WITH AN INFLATABLE PENILE PROSTHESIS (IPP) PRESENTED WITH A DEVICE THAT WAS NO LONGER DEFLATING AND REMAINED IN AN INFLATED STATE FOR OVER ONE WEEK. A REVISION SURGERY WAS PERFORMED, DURING WHICH THE PHYSICIAN CONFIRMED THAT THE RESERVOIR TUBING WAS LINKED AND BOTH CYLINDERS HAD HERNIATED. THE DEVICE WAS EXPLANTED AND REPLACED. THERE WERE NO PATIENT COMPLICATIONS.