inforMED
InjuryJCW

CONCEAL

Received Jul 8, 2026 · Event occurred Jun 19, 2026

Report 2124215-2026-36616 · MDR key 25751073

Device

Generic name

Prosthesis, Penis, Inflatable

Model number

720185-01

Catalog number

720185-01

Lot number

0159398006

Product problems

  • Fluid/Blood Leak
  • Mechanical Problem
  • Material Puncture/Hole

Patient

69 YR · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THIS INFLATABLE PENILE PROSTHESIS (IPP) WAS UNABLE TO CYCLE. A SURGICAL PROCEDURE WAS PERFORMED DURING WHICH THE DEVICE WAS EXPLANTED AND REPLACED. UPON EXPLANT THE PHYSICIAN IDENTIFIED A HOLE AT THE JUNCTURE OF THE PUMP TUBING WHERE IT CONNECTED TO THE KINK RESISTANT TUBING (KRT), AND THERE WAS FLUID LEAKAGE. NO ADDITIONAL INFORMATION WAS PROVIDED.