LIGACLIP
Received Jul 8, 2026 · Event occurred May 27, 2026
Report 3005075853-2026-05438 · MDR key 25748857
Device
Generic name
Clip, Implantable
Manufacturer
Ethicon Endo-surgery, Llc.Catalog number
ER320
Lot number
895D00
Product problems
- Failure to Fire
Patient
Not reported
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Additional Manufacturer Narrative
(b)(4).DATE SENT: 7/8/2026.D4: BATCH # A70028.INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO EES FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE.VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE ER320 DEVICE WAS RECEIVED WITH NO APPARENT DAMAGE. IN ADDITION, THE OPENED PACKAGING WAS RETURNED ALONG WITH THE INSTRUMENT. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS SUBJECTED TO FUNCTIONAL TESTING. DURING TESTING, THE TRIGGER COULD NOT BE ACTUATED, AS IT WAS FOUND TO BE JAMMED. TO FURTHER EVALUATE THE INTERNAL COMPONENTS, THE DEVICE WAS DISASSEMBLED. UPON DISASSEMBLY, THE TRIGGER WAS FOUND TO BE BENT, PREVENTING THE CLIPS FROM FEEDING INTO THE JAWS. FURTHERMORE, EIGHTEEN (18) CLIPS WERE FOUND REMAINING WITHIN THE CLIP TRACK.AS PART OF EES QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS.NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. DEVICE HISTORY REVIEW:A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE BATCH A70028 NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED.THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.