inforMED
InjuryJWH

GMK SPHERE TOTAL KNEE SYSTEM

Received Jul 8, 2026 · Event occurred Jun 1, 2026

Report 3005180920-2026-00671 · MDR key 25747723

Device

Generic name

Gmk-sphere Tibial Insert E-cross - Flex 2r - 10mm

Model number

02.12.E0210FR

Lot number

2345738

Product problems

  • Device Appears to Trigger Rejection

Patient

80 YR · Female

  • Medical device site infection

Narrative

Description of Event or Problem

AT ABOUT 8 MONTHS THE PATIENT CAME IN DUE TO SIGNS OF INFECTION AND THE CAUSE IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE FLEX 2R - 10MM INSERT TO A SAME SIZE INSERT. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Additional Manufacturer Narrative

BATCH REVIEW PERFORMED ON 22 JUNE 2026 GMK-SPHERIKA 02.12.E0210FR GMK-SPHERE TIBIAL INSERT E-CROSS - FLEX 2R - 10MM (K202022) LOT. 2345738: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 18-DEC-2023. EXPIRATION DATE: 2028-12-02. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENTS DURING THE PERIOD OF REVIEW. ROOT CAUSE:INFECTION IS A KNOWN POSSIBLE COMPLICATION OF ANY SURGERY. ALTHOUGH NO ROOT CAUSE CAN BE ESTABLISHED, THERE IS NO INDICATION THAT ANY ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DOCUMENT REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED CAUSE.