inforMED
InjuryLZO

AMISTEM H FEMORAL STEMS

Received Jul 8, 2026 · Event occurred Jun 16, 2026

Report 3005180920-2026-00646 · MDR key 25747720

Device

Generic name

Amistem-h Std. Size 2

Model number

01.18.132

Lot number

132424

Product problems

  • Osseointegration Problem

Patient

NA · Male

  • Inadequate Osseointegration
  • Implant Pain
  • Aseptic loosening

Narrative

Additional Manufacturer Narrative

BATCH REVIEW PERFORMED ON 18 APRIL 2026 STEM: AMISTEM H 01.18.132 AMISTEM-H STD. SIZE 2 (K093944) LOT 2503944: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 02-SEP-2013. EXPIRATION DATE: 31-JUL-2018. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ROOT CAUSE:ASEPTIC LOOSENING IS POSSIBLE LITERATURE DESCRIBED ADVERSE EVENT AFTER PRIMARY JOINT ARTHROPLASTIES AND CAUSES ARE OFTEN UNKNOWN. THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AND THE INVESTIGATION DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED ISSUE OR SYSTEMIC DEFICIENCY.

Description of Event or Problem