inforMED
InjuryJWH

GMK SPHERE TOTAL KNEE SYSTEM

Received Jul 8, 2026 · Event occurred Jun 17, 2026

Report 3005180920-2026-00649 · MDR key 25747719

Device

Generic name

Gmk-sphere Tibial Insert E-cross - Flex 3r - 12mm

Model number

02.12.E0312FR

Lot number

2507478

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

69 YR · Female

  • Joint Laxity
  • Implant Pain

Narrative

Additional Manufacturer Narrative

BATCH REVIEW PERFORMED ON 18 JUN 2026 LOT 2507478: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10-AUG-2025. EXPIRATION DATE: 2030-AUG-26. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ROOT CAUSE:THE SURGEON REVISED LINER HEIGHT, WHICH IS A COMMON PRACTICE TO RESTORE JOINT STABILITY. THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT.

Description of Event or Problem