inforMED
MalfunctionSCV

VERSIUS SURGICAL SYSTEM

Received Jul 8, 2026 · Event occurred Jun 11, 2026

Report 3015488559-2026-00126 · MDR key 25747558

Device

Generic name

Versius Surgeon Console

Catalog number

V-SS-1000

Product problems

  • Circuit Failure

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

CMR SURGICAL LIMITED IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REGULATION 803. THE DEVICE WAS INSPECTED BY THE DISTRIBUTOR. CMR SURGICAL LTD DOES NOT CONSIDER THIS REPORT AND CONTENT TO BE AN ADMISSION THAT ITS PRODUCT IS DEFECTIVE OR THAT IT HAS CAUSED A DEATH OR SERIOUS INJURY.

Description of Event or Problem

IT WAS ALLEGED THAT THE LEFT CLUTCH BUTTON WAS INTERMITTENTLY NOT RESPONDING TO BUTTON PRESS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH VERSIUS SURGICAL SYSTEM. FOLLOWING THE PROCEDURE, THE CONSOLE WAS INSPECTED AND THE LEFT HANDGRIP WAS REPLACED. TELEMETRY ANALYSIS CONFIRMS A HARDWARE ERROR IN THE LEFT HANDGRIP. NO HARM WAS REPORTED IN RELATION TO THIS EVENT. AT THE TIME OF THIS REPORT, NO FURTHER INFORMATION HAS BEEN PROVIDED.