inforMED
MalfunctionNIG

ELECSYS CA 19-9

Received Jul 8, 2026 · Event occurred Jun 16, 2026

Report 1823260-2026-02742 · MDR key 25746240

Device

Generic name

Test, Carbohydrate Antigen (ca19-9)

Manufacturer

Roche Diagnostics

Catalog number

07027028188

Lot number

939827

Product problems

  • No Apparent Adverse Event
  • Prophylactic removal due to corrective action

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH ELECSYS CA 19-9 ASSAY ON A COBAS E 801 ANALYTICAL UNIT.INITIAL RESULT: 62.9 U/ML.1ST REPEAT RESULT: 57.4 U/ML.2ND REPEAT RESULT: 65.2 U/ML (TESTED AFTER 25% POLYETHYLENE GLYCOL (PEG) TREATMENT).3RD REPEAT RESULT: 37.8 U/ML (TESTED ON ABBOTT ALINITY).4TH REPEAT RESULT: 35 U/ML (TESTED ON SIEMENS ATELLICA).THE CUSTOMER ALLEGED THAT ROCHE RESULTS WERE HIGHER THAN THE REFERENCE VALUES FOR THE NORMAL POPULATION.NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE THE LABORATORY AS THEY DID NOT MATCH THE PATIENT'S CLINICAL DIAGNOSIS.THE CUSTOMER REPORTED THE RESULT OBTAINED FROM ABBOTT ALINITY OUTSIDE THE LABORATORY.

Additional Manufacturer Narrative

THE COBAS E 801 ANALYTICAL UNIT SERIAL NUMBER WAS 4(b)(4).THE QC WAS ACCEPTABLE.THE SAMPLE WAS RECEIVED FOR INVESTIGATION ON (b)(6) 2026.THE INVESTIGATION IS ONGOING.