MalfunctionSCV

VERSIUS SURGICAL SYSTEM

Received Jul 8, 2026 · Event occurred Jun 10, 2026

Report 3015488559-2026-00108 · MDR key 25745205

Device

Generic name

Versius Surgeon Console

Catalog number

V-SS-1000

Product problems

  • Connection Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

CMR SURGICAL LIMITED IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REGULATIONS. CMR SURGICAL LTD, DOES NOT CONSIDER THIS REPORT AND CONTENT TO BE AN ADMISSION THAT ITS PRODUCT IS DEFECTIVE OR THAT IT HAS CAUSED A DEATH OR SERIOUS INJURY.

Description of Event or Problem

THE REPORTER ALLEGED THAT DURING SURGICAL SET UP THERE WAS AN ISSUE WITH THE P-STOP. THE BUTTON WAS ACTIVATING WITHOUT IT BEING PRESSED. INSPECTION OF THE DEVICE CARRIED OUT BY A FIELD SERVICE ENGINEER, REVEALED A STICKY RESIDUE COVERING THE BOARD, INCLUDING THE BUTTONS. NO HARM HAS OCCURED TO PATIENT OR USER. AT THE TIME OF THIS REPORT, NO FURTHER INFORMATION HAS BEEN PROVIDED. CMR SURGICAL LTD DOES NOT CONSIDER THIS REPORT AND CONTENT TO BE AN ADMISSION THAT ITS PRODUCT IS DEFECTIVE OR THAT IT HAS CAUSED A DEATH OR SERIOUS INJURY.