inforMED
InjuryKDI

AK 98 MACHINES

Received Jul 7, 2026 · Event occurred Jun 5, 2026

Report 9616026-2026-00091 · MDR key 25736737

Device

Generic name

Dialyzer, High Permeability With Or Without Sealed Dialysate System

Model number

NA

Catalog number

115251

Lot number

NA

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

NA · Male

  • Dyspnea
  • Hemolysis
  • Nausea
  • Vomiting
  • Abdominal Cramps

Narrative

Description of Event or Problem

IT WAS REPORTED THAT 3 HOURS AND 45 MINUTES AFTER DIALYSIS WAS INITIATED USING AN AK 98 MACHINE, A HOME PATIENT EXPERIENCED HEMOLYSIS DUE TO A KINK ON THE ARTERIAL BLOOD LINE ABOVE THE DIALYZER. THERE WERE NO ALARMS REPORTED. THE PATIENT PRESENTED WITH CHEST/ ABDOMINAL CRAMPS, NAUSEA, VOMITING AND DYSPNEA. ADDITIONALLY, IT WAS REPORTED THAT THERE WAS "A SIGNIFICANT DECREASE IN THE HEMOGLOBIN". THE PATIENT WAS TRANSFERRED TO INTENSIVE CARE. IT WAS REPORTED THAT THE PATIENT WAS "RETRANSFUSED". NO ADDITIONAL INFORMATION IS AVAILABLE.

Additional Manufacturer Narrative

E1: INITIAL REPORTER PHONE NO.: (B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.