inforMED
InjuryKDI

FX CORAL FX80 HF 24/CS 1.8SA STEAM

Received Jul 7, 2026 · Event occurred Jun 25, 2026

Report 3002807005-2026-00053 · MDR key 25736499

Device

Generic name

Dialyzer, High Permeability With Or Without Sealed Dialysate System

Manufacturer

St. Wendel Ag

Catalog number

F00012970

Lot number

H8FA17100

Product problems

  • Fluid/Blood Leak

Patient

Not reported

  • Hemorrhage/Blood Loss/Bleeding

Narrative

Additional Manufacturer Narrative

THE PLANT INVESTIGATION IS IN PROCESS. A SUPPLEMENTAL MDR WILL BE SUBMITTED UPON COMPLETION OF THIS ACTIVITY.

Description of Event or Problem

A NURSE REPORTED A DIALYZER BLOOD LEAK DURING A HEMODIALYSIS (HD) TREATMENT. IN A FOLLOW UP, THE NURSE SAID THE LEAK OCCURRED AT THE END OF THE HEMODIALYSIS (HD) TREATMENT. THE LEAK WAS DESCRIBED AS BEING AN INTERNAL LEAK AND WAS SEEN IN THE DIALYSATE CONNECTOR WASTE LINE WHERE FLUID WAS PINK. BLOOD TESTS STRIPS WERE USED AND TESTED POSITIVE FOR THE PRESENCE OF BLOOD. THE 5008X DIALYSIS MACHINE DID ALARM WITH A BLOOD LEAK ALARM. FRESENIUS BLOODLINES WERE BEING USED.THERE WAS NO DAMAGE NOTED ON THE DIALYZER. THERE WAS PATIENT BLOOD LOSS OF 200ML. THERE WAS NO PATIENT INJURY, ADVERSE EVENT OR MEDICAL INTERVENTION REPORTED. THE PATIENT WAS FINISHED WITH TREATMENT.THE DEVICE IS NOT AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION.