inforMED
MalfunctionGCM

VIDEO-OPTIK "ENDOEYE 3D", 30°

Received Jul 7, 2026 · Event occurred Jun 9, 2026

Report 9610773-2026-04765 · MDR key 25734824

Device

Generic name

Rigid Video Laparoscope

Model number

WA50082A

Product problems

  • Poor Quality Image

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE CUSTOMER RETURNED THE RIGID VIDEO LAPAROSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, TECHNICAL ASSISTANCE IDENTIFIED AN IMAGE FAILURE DURING VISUAL INSPECTION. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: FAILURE OF THE VISUAL INSPECTION OF IMAGE OCCURRED DUE TO COMPONENT FAILURE OF CHARGED COUPLED DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.