HU-FRIEDY MFG CO., LLC.
Received Jul 7, 2026 · Event occurred Jun 8, 2026
Report 1416605-2026-00014 · MDR key 25734662
Device
Generic name
#7/8 After Five Gracey Curette Ss
Manufacturer
Hu-friedy Mfg. Co., LlcModel number
SRPG7/86Catalog number
SRPG7/86
Lot number
1023
Product problems
- Break
Patient
NA · Unknown
- Foreign Body In Patient
Narrative
Additional Manufacturer Narrative
THE FAILURE IS MOST LIKELY ASSOCIATED WITH CRACK DEVELOPMENT AND PROPAGATION DUE TO MECHANICAL FATIGUE AND/OR STRESS ACCUMULATED DURING USE. THE OBSERVED CUTTING EDGE DULLING AND LACK OF TIMELY RESHARPENING MAY HAVE CONTRIBUTED TO INCREASED FORCE APPLIED DURING CLINICAL USE, FURTHER PROMOTING CRACK PROPAGATION AND EVENTUAL FRACTURE. NO INDICATION OF MANUFACTURING DEFECT WAS IDENTIFIED.
Description of Event or Problem
DEALER REPORTED THAT CUSTOMER STATED THE INSTRUMENT BROKE OFF WITHIN THE PATIENTS MOUTH. THE BROKEN PART COULD NOT BE REMOVED AND PATIENT NEEDED AN X-RAY AT THE ORAL SURGEON. THE BROKEN PIECE WAS VISIBLE ON THE X-RAY.