inforMED
InjuryKYJ

TOUCH® CMC 1

Received Jul 7, 2026 · Event occurred Jun 9, 2026

Report 3009540749-2026-00005 · MDR key 25733794

Device

Generic name

Finger Prosthesis

Manufacturer

Keri Medical Sa

Model number

NTO08

Lot number

9359

Product problems

  • Inadequacy of Device Shape and/or Size

Patient

52 YR · Female

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

Narrative

Additional Manufacturer Narrative

ALL INFORMATION REASONABLY KNOWN BY THE IMPORTER IS INCLUDED IN THIS REPORT.

Description of Event or Problem

PT RETURNED FOR 2 WEEK POST-OP VISIT AND WAS DISLOCATED. DR NICOSON STATED THAT HE FELT LIKE THE ORIGINAL NECK (08) WAS TOO TIGHT AND CAUSED A VOLAR DISLOCATION. PT TAKEN BACK TO SURGERY ON (B)(6). ORIGINAL NECK (NTO08) WAS REPLACED WITH A (NTO06).