inforMED
MalfunctionOUG

AQURE

Received Jul 7, 2026 · Event occurred Jun 22, 2026

Report 3002807968-2026-00071 · MDR key 25732100

Device

Generic name

Aqure

Model number

933-599

Catalog number

933-599

Product problems

  • Date/Time-Related Software Problem
  • Communication or Transmission Problem

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

RADIOMETER COMPLAINT REFERENCE: (b)(4).ACCORDING TO COMPLAINT, THE CUSTOMER REPORTED THAT WHEN USING "REPROCESS" ON AQURE (SERIAL NUMBER (b)(6)), THE ANALYSIS TIME IS CHANGED TO 00:00 (12:00 AM).THIS INCORRECT TIMESTAMP IS THEN TRANSMITTED TO THE LIS (LABORATORY INFORMATION SYSTEM), WHEREAS THE ORIGINAL ANALYSIS TIME (E.G., 16:21 ON (b)(6) 2026) IS LOST.THE ISSUE IS ALSO OBSERVED FOR BLOOD GAS RESULTS AFTER MODIFICATION AND WAS NOT SEEN PREVIOUSLY.