inforMED
MalfunctionCGX

CREATININE JAFFE GEN.2

Received Jul 7, 2026 · Event occurred Jun 7, 2026

Report 1823260-2026-02717 · MDR key 25731514

Device

Generic name

Alkaline Picrate, Colorimetry, Creatinine

Manufacturer

Roche Diagnostics

Catalog number

06407137214

Lot number

ASKU

Product problems

  • False Positive Result
  • Incorrect, Inadequate or Imprecise Result or Readings
  • Output Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE INITIAL REPORTER RECEIVED A QUESTIONABLE CREATININE JAFFE GEN.2 RESULT FOR ONE PATIENT SAMPLE TESTED ON THE COBAS 8000 C702 MODULE. ON (B)(6) 2026, THE INITIAL RESULT WAS 187 UMOL/L. ON (B)(6) 2026, THE REPEAT RESULT WAS 80 UMOL/L.

Additional Manufacturer Narrative

THE COBAS 8000 C702 MODULE SERIAL NUMBER IS (B)(6). THE FIELD SERVICE ENGINEER INSPECTED THE MODULE, REPLACED THE SAMPLE AND REAGENT PROBES, FLUSHED THE RINSE TUBING, CLEANED THE ULTRASONIC MIXER AND WATER TANK, REPLACED THE CUVETTES, AND VERIFIED THE LEVEL.