inforMED
MalfunctionLIT

ATHLETIS?

Received Jul 6, 2026 · Event occurred Jun 11, 2026

Report 2124215-2026-36115 · MDR key 25728052

Device

Generic name

Catheter, Angioplasty, Peripheral, Transluminal

Model number

H74939347060450

Catalog number

H74939347060450

Lot number

0033997247

Product problems

  • Material Rupture

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT FOREARM SHUNT VEIN SIDE. A 6.0MM X 40MM X 50CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 15 ATMOSPHERES FOR 10 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.