inforMED
InjuryFOZ

SLIDING SAFETY IV CATHETER

Received Jul 6, 2026 · Event occurred Jun 24, 2026

Report MW5190311 · MDR key 25722172

Device

Generic name

Catheter, Intravascular, Therapeutic, Short-term Less Than 30 Days

Manufacturer

Medsource Labs Llc

Lot number

50841/0586

Patient

81 YR · Male

  • Foreign Body In Patient

Narrative

Description of Event or Problem

WHEN STARTING THE PATIENT'S INTRAVENOUS THERE WAS BRISK BLOOD RETURN NOTED; CATHETER WAS ADVANCED WITHOUT DIFFICULTY. WHEN FLUSHING THE PATIENT'S INTRAVENOUS IT BEGAN TO LEAK. I ATTEMPTED TO TIGHTEN THE CATHETER AND IT CONTINUED TO LEAK WHILE FLUSHING. UPON ASSESSING THE INTRAVENOUS SITE I NOTED THAT THE CATHETER WAS NO LONGER INTACT WITH THE HUB. I NOTIFIED NURSE MANAGER AND WE BOTH ASSESSED THE SITE AND NOTIFIED PHYSICIAN. MD IN TO ASSESS THE PATIENT. PER MD HE COULD PALPATE THE CATHETER UNDER THE SKIN IN THE RIGHT WRIST AREA WHERE THE INTRAVENOUS WAS STARTED. EMERGENCY DEPARTMENT MADE AWARE AND PATIENT WILL BE STOPPING BY AFTER INFUSION. THE AREA WAS CIRCLED WITH A MARKER AND EXTRA INTRAVENOUS CATHETER WAS SENT WITH PATIENT FOR A REFERENCE FOR EMERGENCY DEPARTMENT. ALL LOTS OF SAME CATHETER (MCKESSON) WERE REMOVED FROM MED CART. NURSE MANAGER KEPT ALL DEFECTIVE PRODUCT. NURSE MANAGER CALLED EMERGENCY ROOM ON (B)(6) 2026 AND SPOKE WITH PHYSICIAN ASSISTANT THAT TREATED PATIENT AND THEY WERE ABLE TO RETRIEVE AND REMOVE PLASTIC CATHETER SHEATH FROM UNDERNEATH PATIENT'S SKIN AND VEIN. SUTURES WERE PLACED AND ANTIBIOTIC WAS STARTED. TOLD TO FOLLOW UP IN 5 TO 7 DAYS. PATIENT DISCHARGED HOME.