5008X CARESYSTEM + CLIC + CDX
Received Jul 6, 2026 · Event occurred Jun 5, 2026
Report 0001225714-2026-00006 · MDR key 25719904
Device
Generic name
Dialyzer, High Permeability With Or Without Sealed Dialysate System
Manufacturer
SchweinfurtCatalog number
M204441
Product problems
- Adverse Event Without Identified Device or Use Problem
Patient
83 YR · Female
- Low Blood Pressure/ Hypotension
- Syncope/Fainting
Narrative
Description of Event or Problem
IT WAS REPORTED THAT A PATIENT BECAME UNRESPONSIVE DURING HEMODIALYSIS TREATMENT UTILIZING THE FRESENIUS 5008 X HEMODIALYSIS MACHINE. THE PATIENT WAS REVIVED AND THAT PER AUTOMATIC EXTERNAL DEFIBRILLATOR (AED) NO SHOCK WAS ADVISED. ADDITIONAL INFORMATION WAS OBTAINED THROUGH FOLLOW-UP WITH THE CHARGE HEMODIALYIS NURSE AND THE BIOMEDICAL TECHNICIAN AT THE USER FACILITY. PER THE NURSE, ON (B)(6) 2026, THE PATIENT ARRIVED AT CLINIC FOR THEIR REGULARLY SCHEDULED HEMODIALYSIS TREATMENT. THE PATIENT HAS A PRESCRIBED DIALYSIS TREATMENT TIME OF 3 HOURS AND 45 MINUTES AND A PRESCRIBED DRY WEIGHT OF 57 KG. THE PATIENT¿S PRE-DIALYSIS VITAL SIGNS INCLUDED BLOOD PRESSURE OF 163/69 MM/HG, PULSE OF 69 AND PRE-DIALYSIS WEIGHT OF 59.6KG. IT WAS REPORTED THAT THE FRESENIUS 5008X MACHINE PASSED ALL PRE-TREATMENTS TESTING WITHOUT ANY INDICATION OF ANY MACHINE ISSUES. PER THE NURSE, THE PATIENT¿S TREATMENT WAS INITIATED WITHOUT ANY ISSUES. HOWEVER, APPROXIMATELY 3 HOURS AND 28 MINUTES INTO THE HEMODIALYSIS TREATMENT THE PATIENT HAD A HYPOTENSIVE EVENT. IT WAS CONFIRMED THAT THE PATIENT¿S BLOOD PRESSURE DECREASED TO 77/40 MM/HG WITH A PULSE OF 94 WITH A LOSS OF CONSCIOUSNESS (SYNCOPE). AS A RESULT, THE PATIENT¿S BLOOD WAS RETURNED AND THE TREATMENT WAS DISCONTINUED. FURTHERMORE, THE PATIENT WAS GIVEN OXYGEN VIA NASAL CANNULA (4 LITERS) AND RECEIVED APPROXIMATELY 700 ML OF NORMAL SALINE. THE PATIENT REPORTEDLY HAD SPONTANEOUS RETURN OF CONSCIOUSNESS WITHOUT NEED FOR FURTHER INTERVENTION. POST-EVENT VITAL SIGNS INCLUDED A BLOOD PRESSURE OF 179/73 MM/HG WITH A PULSE OF 62 AND POST DIALYSIS WEIGHT OF 59.6KG (UNCHANGED). THE NURSE INDICATED THAT THE PATIENT WAS SENT TO THE HOSPITAL VIA AMBULANCE FOR FURTHER MEDICAL EVALUATION. HOWEVER, THE NURSE STATED THE PATIENT WAS DISCHARGED FROM THE HOSPITAL WITHIN A FEW HOURS (NOT ADMITTED) WITHOUT REQUIRING ANY FURTHER MEDICAL INTERVENTION. ON (B)(6) 2026, THE PATIENT RESUMED THEIR NEXT SCHEDULED TREATMENT WITHOUT ANY FURTHER REPORTED ISSUES. THE NURSE STATED THE PATIENT EVENT OF HYPOTENSION IS NOT SUSPECTED TO BE CAUSED BY THE 5008X MACHINE. THE NURSE STATED THAT HYPOTENSION IS A COMMON COMPLICATION DURING HEMODIALYSIS DUE TO FLUID REMOVAL FROM THE VASCULATURE. HOWEVER, PER CLINIC PROTOCOL, THE MACHINE WAS PULLED POST EVENT FOR TESTING AFTER THE ADVERSE EVENT OCCURRED. THE BIOMEDICAL TECHNICIAN CONFIRMED THAT THE 5008X MACHINE WAS PULLED FROM SERVICE POST PATIENT EVENT, PER CLINIC PROTOCOL. THE BIOMEDICAL TECHNICIAN INDICATED THAT THE MACHINE PASSED ALL TESTS POST EVENT AND THAT THERE WERE NO MACHINE MALFUNCTIONS OR ISSUES RELATED TO THE EVENT. IT WAS CONFIRMED THROUGH DOCUMENTATION ON THE QUALITY REPORT FOR POST EVENT TESTING OF THE 5008X MACHINE THAT THE MACHINE PASSED ALL FUNCTIONAL TESTS AND ALARMED APPROPRIATELY WHILE IN DIALYSIS MODE. THE BIOMEDICAL TECHNICIAN REPORTED THAT THE 5008X MACHINE WAS RETURNED BACK INTO SERVICE.