THORATEC HEARTMATE 3 LVAD IMPLANT KIT
Received Jul 6, 2026 · Event occurred May 12, 2026
Report MW5190281 · MDR key 25719466
Device
Generic name
Ventricular (assist) Bypass
Manufacturer
Abbott Laboratories / Thoratec Corp.Patient
47 YR · Male
- Hemorrhage/Blood Loss/Bleeding
Narrative
Description of Event or Problem
AFTER INITIAL IMPLANT OF HEARTMATE 3 DEVICE, IT WAS DISCOVERED THAT THE PUMP WAS NOT CORRECTLY LOCKING INTO THE APICAL SEWING RING. THE PUMP HEAD APPEARED TO BE LOOSE DESPITE BEING FULLY LATCHED ON AND IT WOULD EASILY TWIST AROUND WHILE FULLY LATCHED ALONG ITS AXIS WHILE ATTACHED TO THE SEWING RING. BOTH THE OUTFLOW GRAFT AND INFLOW SIDE SEEMED TO BE DEFECTIVE, AND WOULD NOT SECURELY TIGHTEN DOWN AS INTENDED. THIS CAUSED THE OUTFLOW GRAFT TO BECOME DISCONNECTED DURING NORMAL HANDLING, WITH SIGNIFICANT BLEEDING AROUND THE CANNULA. THE DEVICE WAS REMOVED, THE PATIENT HAD TO BE PLACED BACK ON BYPASS, AND A NEW HEARTMATE 3 DEVICE WAS OPENED AND IMPLANTED.