inforMED
MalfunctionSCV

HUGO¿ RAS SYSTEM

Received Jul 6, 2026 · Event occurred Jul 2, 2026

Report 1219930-2026-04274 · MDR key 25719341

Device

Generic name

Modular Electromechanical Surgical System

Manufacturer

Surgical

Model number

MRASC0005

Catalog number

MRASC0005

Product problems

  • Use of Device Problem
  • Complete Loss of Power
  • No Tactile Prompts/Feedback

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT, PRE-OPERATIVELY, THE AC POWER CABLES WERE DISCONNECTED FROM THE UNIT. AFTER POWER REMOVAL, THE LOWER BATTERY REMAINED ACTIVE AND CONTINUED TO DISCHARGE WHILE THE SYSTEM WAS UNPLUGGED. THE BATTERY LEVEL DROPPED TO APPROXIMATELY 70 PERCENT. THE UNIT WAS SET TO RECHARGE; HOWEVER, DURING THE RECHARGE PERIOD, THE POWER BUTTON (LIGHT EMITTING DIODE) LED WAS NOT OBSERVED PULSING. AFTER THE SYSTEM RECHARGED TO 100%, THE SYSTEM WAS MANUALLY REBOOTED AND OPERATED PROPERLY. NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.